Description
The FF-8610 is a polarization analysis device for measuring spatial collimation light and fiber optic sources. The instrument uses the rotating wave plate method to measure the degree of polarization. The maximum sampling rate is 20hz. The FF-8610 is equipped with software that displays the polarization state in real time on the Bonga Sphere window. The polarization extinction ratio (PER) of narrow-band light can be measured through the measurement technology of heating optical fiber by combining with the polarization state heat cycle heater.
Application
1. Polarization state analysis and monitoring
2. Polarization measurement
3. Extinction ratio measurement (with fiber self-circulating heater)
4. Rotation Angle measurement
| Model | FF-8610 |
| Wavelength range | 1270-1650nm |
| SOP accuracy | 0.25° |
| DOP accuracy | 1% |
| PER measuring range | 0-40dB |
| PER resolution | 0.1dB |
| Input optical power range | -40~+10dBm |
| Sampling rate | 100hz |
| Port | FC or space collimation light |
| Light aperture | 3mm |
| Maximum deviation Angle of collimating light | 2° |
| Size | 235*104*300 mm |
FAQ
Q1: What is the main measurement principle and primary application of the FF-8610 SOP Rotary Analyzer?
A: The FF-8610 is a precision optical instrument engineered to measure and monitor polarization characteristics for both fiber optic light sources and spatial collimated light beams. Utilizing the rotating waveplate method, it determines the State of Polarization (SOP), Degree of Polarization (DOP), Polarization Extinction Ratio (PER), and optical rotation angle.
Q2: What key technical features distinguish the FF-8610 analyzer?
A: Key operational features include:
- Rotating Waveplate Method: Accurately measures polarization states with high repeatability.
- Flexible Input Port: Accommodates both FC fiber connectors and free-space collimated light with a 3 mm aperture and max deviation angle of 2°.
- Real-Time Poincaré Sphere Display: Dedicated software visualizes polarization trajectories and parameters dynamically on a Poincaré sphere window.
- Narrow-Band PER Testing: Measures narrow-band light extinction ratios when combined with a polarization heat cycle heater.
Q3: What are the main optical performance specifications of the FF-8610?
A: Core performance metrics include:
- Wavelength Range: 1270 nm to 1650 nm
- SOP Accuracy: 0.25°
- DOP Accuracy: 1%
- PER Range & Resolution: 0 dB to 40 dB range with 0.1 dB resolution
- Input Optical Power: -40 dBm to +10 dBm
- Sampling Rate: Up to 100 Hz
- Chassis Dimensions: 235 × 104 × 300 mm
Q4: Who manufactures the FF-8610 Analyzer and what is its reference price?
A: Manufactured by FirstFiber Technologies, the FF-8610 SOP Rotary State Of Polarization Analyzer carries a reference price of $5,000 USD.
Q5: How can I request volume pricing or schedule a product demonstration?
A: You can click “Add to Quote” on the product page for custom volume pricing inquiries or select “Request a Demo” to submit an evaluation request.